Customization: | Available |
---|---|
Warranty: | 1 year |
Type: | XRF Optical Spectrum Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
DX-2800M Portable Xrf Spectrometer For Gold Purity And Karat Test
Analysis Of Precious Metals And Identification Of Carats
At present, the use of XRF analyzer for chemical composition analysis and purity
determination of precious metals has become a widely used, extremely popular, and
supported by national standards, and reliable methods. The use of XRF for the analysis of
precious metals is a faster and more economical method of multi-element detection than spark
test and chemical reagent testing.
The XRF Analyzer provides users with an easy-to-use, cost-effective inspection technology: no
need to inspect the tool into the material of the sample being tested, or damage the sample.
The chemical composition information of the alloy can be obtained.thus determines the carat
rating of the sample. Whether you are buying precious metals such as gold, selling or producing jewelry, making metals, or recycling used metals, you need a quick way to determine the carat grade and other precious and non-precious metal content with extreme precision.
In the first time, to effectively control product quality, set a reasonable price. The xrf analyzer quickly provides accurate analysis of all types of samples.
Advantages And Characteristics Of Analysis
Instrument Performance And Configuration
Brand | DahoMeter | ||
Model | DX-2800 | DX-2800M | |
Computer Built In | NO | NO | |
Detector Type | Si-pin (USA) | SDD(USA) | |
Multichannel Analyzer | DahoMeter-DMCP | ||
Sample Chamber Size | 380 * 380 * 60 (mm) | ||
Test Time | 5sec ~ 30sec | ||
Measuring Element | 72 elements from sodium (Na) touranium (U) between all the elements | ||
Analysis Range | 1ppm to 99.99% | ||
Accuracy | 0.03% (Au≥96%) | 0.01% (Au≥96%) | |
Sample Form | Solid, powder, liquid | ||
Tube Voltage | 5KV ~ 50KV | ||
High Voltage Power Supply | 0 ~ 50KV | ||
Camera | HD camera | ||
Filters | Selectable customized switches | ||
Analysis Software | Qualitative and quantitative analysis software | ||
External Dimensions | 480 * 480 * 390 (mm) | 480 * 480 * 390 (mm) | |
Weight | 38Kg | 38Kg | |
Instruments Environmental Requirements | |||
Ambient Temperature | -5 ° C ~ 40 ° C (Suggested that in air conditioning room) | ||
Relative Humidity | 15% ~ 90% | ||
Power Requirements | AC 220V ± 5V, 50/ 60 HzNo high power electromagnetic and vibration interference sources nearby |
Cetification