DH-OA-150 Ozone Aging Test Chamber. Ozone Accelerated Weathering Testing Chamber
Specification
1.150L (L stands for liter)
W500 X H600 x D500mm (W represents width, H represents height, D represents depth)
W1100 X H1700 x D1050mm (W represents width, H represents height, D represents depth)
About 220kg
AC220±5%V /50±0.5Hz
1. Application
The content of ozone in the atmosphere is very small but it is the main factor of rubber cracking. This product simulates and strengthens the ozone conditions in the atmosphere, studies the law of the effect of ozone on rubber, and quickly identifies and evaluates the ozone resistance and aging resistance of rubber. The method of anti-ozonant protection efficiency. Reference standard: GB/T7762-2003 vulcanized rubber or thermoplastic rubber anti-ozone cracking static tensile test method, GB/T2951.21-2008 cable insulation and sheath material general test method, GB/T 11206-2009 Rubber aging test surface cracking method, and other test methods specified in related standards.
Principles and methods
Test methods can be divided into "static" and "dynamic" according to test methods and standards. Static means that the sample is stretched and placed in the test box for testing, while dynamic means that the sample is placed in the test box. On the fixture, do the test while stretching, and the fluctuation of the stretching is about 5%-45% of the tested sample itself
2. Main technical Parameters
2.1. Temperature range: 0ºC-80ºC
2.2. Temperature fluctuation: ±0.5ºC Temperature distribution accuracy: ±2.0ºC
2.3. Ozone concentration accuracy: 10~400pphm ±10%
2.4. Raising frequency of sample rack: 2R/min (optional)
2.5. Air velocity: 12~16mm/s
2.6. Sample rack speed: 360 degree rotating sample rack (rotating speed 1~3 revolutions per minute or optional adjustable)
3.Structural Characteristics
3.1. Structure: One machine.
a) 3.2. Material-inner box: SUS304 stainless steel plate
b) Outer box: SUS304 stainless steel plate or baking varnish (beige white).
c) Thermal insulation: dual protection of rigid polyurethane foam + aluminum silicate thermal insulation cotton.
a) 3.3. Box door-single open door, left open.
b) Two silicone rubber sealing strips for window frame/door frame and anti-condensation electric heating device.
c) Observation window W300xH450mm and lighting lamps are equipped on the doors and windows.
3.4. Requirements of the sample rack
a) (Dynamic experiment; optional)-Fix both ends of the sample with clamps under the required elongation. When in contact with ozonized air, the length of the sample should be basically parallel to the direction of airflow
b) The fixture should be made of a material (such as aluminum) that does not easily decompose ozone.
c) Make the sample rotation speed between (20^-25) mm/s, in a vertical dry air plane, each sample continuously moves along the same path
d) The time for one rotation of the same sample is (8^-12) min. The scanning area of the sample is at least 40% of the effective area of the test box.
e) The sample preparation should meet the requirements of GB/T 9865.1. The sample is best cut from the newly molded test piece, if necessary, it can be cut from the finished product. The sample should be at least 3 samples per test condition. The width of the long standard sample is not less than 10mm, the thickness is 2.0mm±0.2mm, and the length of the sample between the two ends of the clamp before stretching is not less than 40mm.
Standard static test fixture-
- a) 3.5. Control Panel-Controller Display
- b) Power switch
- c) Illumination button,
- a) 3.6. Cabinet configuration-a stainless steel SUS #304 rack turntable inside
- b) φ50mm lead hole lead hole: 1 piece, equipped with hole cover and silicone rubber plug.
- c) 4 movable casters
- d) 4 horizontal fixed tripods
- There is a drainage hole at the bottom of the box body to facilitate the drainage of the condensate water of the unit;
4.air circulation adjustment
4.1. Adjustment method
4.2. Circulating fan-FLOW THROW air supply mode; horizontal diffusion and vertical heat exchange arc circulation.
Power-saving motor + stainless steel extension shaft is connected with multi-centrifugal circulating fan.
5. heating system
5.1. Heater
5.2. Control method-fin-type radiator tube-shaped stainless steel electric heater heating.
-ID adjustment, actuator: solid state relay.
6.refrigeration system
6.1. Working method: mechanical refrigeration/air-cooled.
6.2. Control mode: The controller of the control system automatically adjusts the operating conditions of the refrigerator according to the test conditions.
6.3. Compressor: a well-known European and American brand low-temperature high-efficiency fully enclosed compressor. (France; Taikang)
6.4. Evaporator: fin type multi-stage automatic load capacity adjustment. (Japan; Saginomiya)
6.5. Condenser: wind type condenser. (Japan; Shuangling)
6.6. Expansion system: refrigeration system with automatic capacity control
6.7. Refrigerant: environmentally friendly refrigerant R404 (the ozone depletion index is 0). (United States; DuPont)
6.8. Refrigeration process: Nitrogen protection welding, two-stage rotary vane pump vacuum to ensure that the interior of the refrigeration system is clean, the bottom of the compressor is designed with a water tray, and the condensed water is drained out of the box through the drain pipe at the rear of the box.
a)6.9. Refrigeration accessories: expansion valve (Denmark; DANFOS S)
b) Solenoid valve (Japan; Omron)
c) Filter (Denmark; DANFOS S)
d) Pressure controller (Denmark; DANFOS S)
e) Globe valve (Italy; CASTEL)
f) Oil separator (Europe and America; ALCO)
6.10. Accessories: desiccant, oil separator, refrigerant flow window, repair valve, high pressure protection switch. .
7. Ozone detection system
7.1 Ozone generator-(Using voltage silent discharge tube to generate ozone)
7.2 Ozone sensor-ozone concentration sensor imported from the UK, which can realize precise control
7.3 Ozone exhaust system-After the experiment is completed, the ozone in the experiment box is exhausted to the outdoors through a dedicated exhaust fan and air duct
7.4 Ozone flow regulator-using a high-precision glass flow rotor meter for precise adjustment
8. Control system
8.1. Sensor: (Using voltage silent discharge tube to generate ozone)
8.2. Controller: British imported ozone concentration sensor, which can realize precise control
The temperature sensor adopts platinum DIN PT-100Ω, imported from Korea TEMI, 7-inch color LCD touch-sensitive human-machine interface controller, large field of view screen, and screen lock (LOCK) function. Chinese and English switching
8.3. Controller specification-accuracy: ±0.1ºC+1digit.
a) Temperature resolution: 0.1.
b) Setting range of each time: 0~530 hours can be set freely.
c) Temperature input signal Platinum PT100 type.
d) P.I.D control parameter setting, P.I.D automatic calculation.
a) 7.4. Screen display function-using the graphic control software screen dialogue mode, the screen directly touches the options: including program setting, curve display, historical data, manual operation, automatic operation, auxiliary setting, etc.
b) Display execution status, temperature and time setting value, remaining time, remaining cycle times.
c) With a separate program editing screen, temperature, time and cycle times can be input.
d) The temperature program setting is displayed as a graph curve, with real-time display program curve execution function.
e) Display the fault status and explain the troubleshooting method.
f) The temperature setting (SV) and actual (PV) values are directly displayed.
g) The screen can be adjusted for backlight, and the screen display protection function can be used for timing and closing settings.
8.5. Program capacity and
a) Control function-The production of the program adopts the dialogue function.
b) Cycle times: up to 1-9999 times adjustable.
c) Available program groups: 50 PATTEN at most (that is, 50 test specifications can be independently set and stored)
d) With power-off program memory, it will automatically start and continue to execute the program after power is restored.
e) The graph curve can be displayed in real time when the program is executed.
f) With scheduled startup and shutdown functions.
g) With date and time adjustment functions.
9. safety protection device
a) 9.1. Refrigeration protection-compressor overheating protection switch
b) Compressor high pressure protection switch
c) Compressor overcurrent protection switch
a) 9.2. Heating protection-heater surface temperature over-temperature protection
b) Heater short circuit and overload protection device
a) 9.3. Other protections-circulating motor overheating protection
b) Condensing fan overheat protection
c) High temperature, low temperature, over temperature protection
d) Reverse phase and phase loss protection device,
e) No fuse switch overload protection.
f) Power-off protection device for smoke alarm.
g) Line fuse and fully sheathed terminal
10. Accessories and random information
10.1. Shelf
10.2. Shelf track
10.3. Power-on test hole
10.4. Power cord
10.5. Drain
10.6. Air supply interface
10.7. Moving casters
10.8. Fixed foot cup
10.9. Maintenance Manual
10.10. Controller specification
10.11. Warranty card-SUS304 stainless steel shelf 2 layers (single layer load 10kg)
4 stainless steel shelf rails with adjustable pitch 40mm
1 power-on test hole, located on the left side of the box, with cover and soft plug, size ¢50mm×100mm
One power cord (length 2 meters)
One drainage hole
One air source interface
4 movable casters
4 horizontal fixed foot cups
An operation and maintenance manual
A controller specification
A warranty card
11. Equipment use conditions
11.1. Power
11.2. Environmental conditions
11.3. Channel requirements
-The power cord is at the back of the machine/AC220±5%V /50±0.5Hz
Temperature: 5ºC~30ºC Relative humidity: ≤85%
This device is heavy. Please consider the installation site aisles and handling methods in advance (see the main technical indicators for external dimensions and weight)
a) Site requirements-It is better to have a drainage floor drain or connect a drainage pipe near the test box.
b) Place the on-site wall to configure the equipment with corresponding capacity air or power switches
c) Use the ground site to be level, well ventilated, free of flammable, explosive and corrosive gases and dust
d) Product appearance drawing:
Packaging & Shipping
Company Information
We, Dongguan Hongtuo Instrument Co., Ltd, a China based company, are pleased to introduce ourselves as a notable testing equipments and machines manufacturer.Our company was established in the year 2000 with two primarily motto, which includes complete client satisfaction and to reach top-most position in the international market. The exceptional quality of our products like Gold Tester,Gold Testing Machine,Density Meter,Universal Tensile Testing Machine,Climatic Chamber,Melt Flow Index Tester,Charpy Impact Testing Machine,Tap Density Tester,Abrasion Tester,Ultrasonic Flaw Detector,Surface Roughness Tester and other Laboratory Testing Equipment, etc. is our company's USP.