DH-OI-02 Limiting Oxygen Index Tester Price, Digital Oxygen Index Test Apparatus, Limitation Oxygen Index Test Equipment
Model: DH-OI-02
Standards
Comply with GB/T2406 "Plastic Flammability Test Method-Oxygen Index Method" and JIS7201, BS2782, ANSI/ASTM, D2863, ISO4589, test standards.
Application
It is used to judge the difficulty of burning when the material is in contact with flame in the air. It is suitable for homogeneous solid materials, laminated materials, soft sheets and film materials.
Parameter
Oxygen concentration (oxygen index) | direct reading with digital display |
Oxygen concentration (oxygen index) sensor | imported from the United States. |
Measuring range | 0-80% |
Resolution | 0.1L/min |
Measurement accuracy | 2.5 level |
Combustion cylinder | Quartz glass cylinder, inner diameter:75mm height: 480mm |
Gas flow rate at the top of the combustion tube | 90mm±10mm/s |
Gas flow rate in the combustion cylinder | 40mm±2mm/s |
Input pressure | adjustable from 0.25 to 0.5 Mpa; |
Igniter | dedicated butane handheld electronic electric firearm |
Gas | industrial grade oxygen, nitrogen. |
Timer accuracy | ±0.5s |
Power requirements | AC220 (±10%) V/50HZ. |
The two pressure gauges provide oxygen output pressure and nitrogen output pressure respectively, and the air pressure can be adjusted through a regulating valve. | |
Oxygen and nitrogen mixed gas inlet: including pressure regulator valve, flow regulating valve and mixing chamber. Specimen holder can be used for soft and hard plastics. |
Chassis and part of the structure
4.1. Control box: The square shape is processed by the processing machine tool, and the arc shape is beautiful and generous. The surface is treated with beautiful paint, anti-corrosion and anti-rust.
4.2. Combustion tube: high-temperature high-quality glass tube (inner diameter ¢750mm, length 480mm)
4.3. Outlet inner diameter: φ40mm
4.4. Suspended: Witness can be clamped freely.
Electronic control system
5.1 The relay control system is adopted, and all electronic components are made of domestic brands, which can better ensure the performance and detection accuracy of the machine.
5.2 The equipment has the following safety protections: power supply overload, short circuit protection, and control loop overload.